Symbols, error messages and troubleshooting
13
This error message ... means:

O You bent or moved the test strip dur-
ing the test.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O You inserted a used test strip.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O You used a test strip that is past its
expiry date.
Push the ejector down to eject the
test strip. Test again with test strips
that have not expired. Remove the old
code chip and insert the one belong-
ing to the new test strips.
