Technical data
14
Performance analysis The performance data for the Accu-Chek Go
system (Accu-Chek Go meter with
Accu-Chek Go test strips) were obtained
using capillary blood from diabetic patients
(method comparison, accuracy), venous
blood (repeatability) and control solution
(reproducibility). The system is calibrated
with venous blood containing various levels
of glucose. The reference values are
obtained using the hexokinase method. For
method comparison, the results were com-
pared with results obtained using the hexo-
kinase method with deproteinization (auto-
matic analyzer). The hexokinase method is
traceable to an NIST standard.
The Accu-Chek Go system meets the ISO 15197 requirements.