Accu-Chek Spirit Guide de l'utilisateur Page 145

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 226
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 144
Data transfer
143
Check the configuration on your insulin pump to
make sure that all parameters are written
correctly.
Set up your insulin pump with a cartridge,
adapter and a new infusion set and put it into
RUN, if necessary.
Note Your insulin pump automatically returns to STOP 15 minutes
h after the last data transfer respectively when the DATA
TRANSFER screen has been selected.
You may press d+a simultaneously to exit while data is
being transferred. The data transfer is interrupted and an
error E12: DATA INTERRUPTED occurs.
See section 10.2.11 “Error E12: DATA INTERRUPTED” for
further information.
Vue de la page 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 225 226

Commentaires sur ces manuels

Pas de commentaire