Alerts and errors
161
10.2.9 Error E10: CARTRIDGE ERROR
If the CHANGE THE CARTRIDGE function was not correctly carried out, an
error E10: CARTRIDGE ERROR will occur.
Press f twice to turn off and confirm the error.
Remove or disconnect the infusion set from your
infusion site.
Remove the cartridge. Start the CHANGE THE
CARTRIDGE menu. Reinsert the cartridge.
Prime the infusion set.
Reconnect the infusion set (tubing) at your
infusion site or insert the new infusion set.
When finished, put your insulin pump into RUN.
If an error E10: CARTRIDGE ERROR occurs
during the winding of the piston rod, insert a
new battery into the pump and repeat the steps
above.