Accu-Chek Go Manuel de service Page 156

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 208
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 155
156
Measurement and storage conditions
Farbe: P 021 – Prüfmittelnummer: 201
O Blood glucose tests can be performed at the limits of the permitted
range (between +5 °C and +10 °C or between +40 °C and +45 °C).
If the temperature is within one of these ranges, the thermometer
symbol flashes in the display.
Do not use results obtained at the limits of the permitted
temperature range as a basis for making therapeutic deci-
sions, as they could be incorrect. Incorrect results can cause
the wrong therapeutic decision to be taken and so lead to
serious adverse health effects.
O Tests cannot be performed at temperatures below +5 °C and above
+45 °C.
In this case the following message is displayed:
Vue de la page 155
1 2 ... 151 152 153 154 155 156 157 158 159 160 161 ... 207 208

Commentaires sur ces manuels

Pas de commentaire