Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
This error message ... means:

O The battery is empty or a meter error
has occurred.
Remove the test strip by pushing the
ejector down and re-insert it. If the
error message persists, change the
battery.
O The meter was exposed to a strong
electromagnetic field. Eject the test
strip or turn the meter off. Move
somewhere else or turn off the
source of the field.
Vue de la page 177