Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
If this happens ... it means:

O Condensation has affected the elec-
tronics.
Allow the meter to dry slowly.
O The meter is defective.
Please contact your local customer
support and service centre (see
Chapter 19).
The meter is turned on,
but the display is blank.
The display is defective.
Please contact your local customer sup-
port and service centre (see Chapter 19).
The clock has stopped or
the clock is slow.
The meter was exposed to a temperature
below -10 °C and the battery is begin-
ning to freeze.
Turn the meter off.
Move to a place where the temperature is
between +10 °C and +40 °C and wait for
the meter to warm up.