Accu-Chek Softclix Informations techniques Page 76

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 92
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 75
Chapter 5: Maintenance and Troubleshooting 74 ACCU-CHEK Customer Care Service Center 1-800-858-8072
Display Messages and Troubleshooting (continued)
• Blood was applied to the test strip too soon. Slide and release to turn off the meter
and release the test strip, then turn on the meter again. Repeat the test, waiting for the
flashing test strip and blood drop symbol before touching blood or control solution to
the end of the test strip.
The test strip was bent before the test started. Repeat the test with a new test strip.
The measurement window is dirty. Clean the measurement window area and repeat the
test. See
Chapter 5, Cleaning Your Meter.
There is moisture in the meter. Slide and release to turn off the meter. Let the meter
dry at room temperature.